TELEFLEX MEDICAL WECKVISTA ACCESS BALLOON PORT 12MMX70MM; LAPAROSCOPE, GENERAL & PLASTIC Back to Search Results
Catalog Number 412944
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the balloon got burst during a urological surgery.Therefore, the device was removed and replaced with a new one to complete the procedure.Nothing fell/remained in the patient, and no injury to the patient was reported".The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review could not be performed as no lot number was reported.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that "the balloon got burst during a urological surgery.Therefore, the device was removed and replaced with a new one to complete the procedure.Nothing fell/remained in the patient, and no injury to the patient was reported".The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The reported complaint of "burst - balloon" could not be confirmed based upon the sample received.Upon functional inspection, no leak was observed.The tecate manufacturing site was contacted as part of this investigation.The manufacturing site confirmed that no functional defect could be concluded for the returned sample because the balloon inflated properly.A dhr review was performed with no evidence to suggest a manufacturing related cause.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WECKVISTA ACCESS BALLOON PORT 12MMX70MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18169533
Report Number3003898360-2023-01552
Device Sequence Number7661255
Product Code GCJ
UDI-Device Identifier04026704559763
UDI-Public04026704559763
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/16/2024
Device Catalogue Number412944
Device Lot Number73E2200603
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2023
Supplement Date Received by Manufacturer12/27/2023
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-