BOSTON SCIENTIFIC CORPORATION VISIONIST CRT-P; CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) Back to Search Results
Model Number U225
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Bradycardia (1751)
Event Date 10/24/2023
Event Type  malfunction  
Event Description
It was reported that stored episodes of noise, oversensing, and pacing inhibition were noted on the right ventricular (rv) channel of this cardiac resynchronization therapy pacemaker (crt-p).The health care professional (hcp) reprogrammed the device to doo to avoid rv sensing and requested technical services (ts) analysis.At this time, no further data is available for analysis.Ts provided troubleshooting recommendations.No adverse patient effects were reported, this rv lead remains in-service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18169867
MDR Text Key328535410
Report Number2124215-2023-65275
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/20/2022
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number717499
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/10/2023
Initial Date FDA Received11/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-