Event or Problem Description
It was reported the patient underwent a left knee revision approximately one month post implantation due to instability.During the revision, it was noted that the femoral had fractured.No additional patient consequences were reported.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-03284, 0001822565-2023-03285, and 0001822565-2023-03286.D10 medical devices: distal femoral xt component size b left catalog#: 00585004201; lot#: 65322758; femoral hinge service kit xt size b catalog#: 00585007112; lot#: 65221250.G2 foreign source: south africa.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10 medical devices: distal femoral xt component size b left; catalog#: 00585004201 lot#: 65322758.Femoral hinge service kit xt size b; catalog#: 00585007112 lot#: 65221250.Articular surface with segmental hinge post size b 14 mm height; catalog#: 00585002014 lot#: 66012050.Cerclage cable with crimp 1.8 mm dia.Cable 22 in.(559 mm) length; catalog#: 00223200118 lot#: 65355726.Optip 80 refob plus bone cmt-3; catalog#: 4722502117-3 lot#: ax30bd1106.Optip 80 refob plus bone cmt-3; catalog#: 4722502117-3 lot#: ax30bd1106.Segment with male/female taper 30 mm length; catalog#: 00585004603 lot#: 65492481.Stem collar 25 mm o.D.; catalog#: 00585204025 lot#: 64635296.Proximal tibial component (tibial body x1 tibial bushing x1) size 1; catalog#: 00585000110 lot#: 64953863.Fluted stem extension straight 13 mm diameter 130 mm stem length; catalog#: 00585205013 lot#: 65722452.
 
Event or Problem Description
No further event information available at the time of this report.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned femoral found it to exhibit signs of being implanted and the tabs are fractured off.Medical records and radiographs were provided and reviewed by a health care professional.X-ray reports dictate findings of prosthesis in position without loosening or evidence of complications.A leg length discrepancy was identified.Device history record was reviewed and no discrepancies were found.Review of complaint history found no additional related issues for this item and the reported part and lot combinations.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYETHYLENE INSERT XT SIZE B
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18169997
Report Number0001822565-2023-03285
Device Sequence Number7716661
Product Code KRO
UDI-Device Identifier00889024633889
UDI-Public(01)00889024633889(17)280325(10)65858288
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K150028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00585001296
Device Lot Number65858288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer12/18/2023
04/26/2024
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date01/12/2024
04/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10; SEE H10
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight82 KG
-
-