| Catalog Number |
07729952160 |
| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The meter was requested for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation of a display issue with coaguchek vantus meter serial number (b)(6).The left side of the meter display is faint and the right side is white.The customer is able to navigate the menus and use the meter despite the display issue.The display being faint on one side and white on the other makes the result difficult to read.No misinterpretation of results reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The meter was returned for investigation.The meter was powered on and most of the display stays white.The removed display shows half-round cracks at the bottom edge of the glass in the cable connection area.The device showed no damage which suggests a handling error.Medwatch fields d9 and h3 have been updated.
|
| |
|