• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ACET CUP INTRODUCER 32; JRN : INSERTION DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ACET CUP INTRODUCER 32; JRN : INSERTION DEVICES Back to Search Results
Catalog Number 962636000
Medical Device Problem Codes Contamination /Decontamination Problem (2895); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that introducer has damaged threads.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : broken items due to damaged threads.No surgery delay the device associated with this report was returned to depuy synthes for evaluation.Evaluation of the device can confirm the foreign substance overall the inner sections of the device, no other issue is seen per instructions for use , care should be taken to remove any debris, tissue, or bone fragments that may collect on the instrument.Instruments with cannulas, moving parts, or spring mechanisms require additional cleaning steps to ensure proper removal of all trapped debris.In addition to other important steps to follow, the ifu instructs the use of a soft bristle brush to remove all traces of blood and debris, paying close attention to threads, crevices, seams, and any hard-to-reach areas of the device features.If the instrument has sliding mechanisms or hinged joints, actuate the area to free any trapped blood and debris.Based on the observed unclean condition of the device features, it does not appear the device was properly cleaned as prescribed by the ifu.A dimensional inspection and a functional test were not performed for the acet cup introducer were not performed since it is not applicable for the complaint condition the overall complaint was confirmed as the observed condition of the acet cup introducer would contribute to the complained device issue.Based on the investigation findings, the potential cause is failure to follow instructions.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACET CUP INTRODUCER 32
Common Device Name
JRN : INSERTION DEVICES
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18170332
Report Number1818910-2023-23604
Device Sequence Number7597877
Product Code HWR
UDI-Device Identifier10603295243588
UDI-Public10603295243588
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number962636000
Device Lot Number0SA246653
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2023
Supplement Date Received by Manufacturer12/05/2023
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date12/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/07/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-