| Catalog Number |
03333701190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).The calibration trace data provided was acceptable.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable alp2 alkaline phosphatase acc.To ifcc gen.2 results for 2 patient samples on a cobas 6000 c501 module.This medwatch will cover alp2.Refer to medwatch with a1 patient identifier (b)(6) for information on the ldh results.On (b)(6) 2023, sample 1 had an initial alp result of 802 u/l.Th sample was repeated twice and the results were 467 u/l and 364 u/l.On (b)(6) 2023, the initial result was 355.8 u/l.The repeat result was 143.9 u/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined the issue was consistent with a high leukocyte concentration in the samples leading to an accumulation of cells close to the plasma surface.Falsely high results can occur if blood cells or cell debris are not properly removed in the pre-analytical sampling process.The most likely root cause is a known interference caused by a pre-analytical issue.Correct pre-analytic sample handling is within the customer's responsibility.There is no evidence to suggest a malfunction of the device.
|
| |
|