• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY UNI TIBINRT S5-6RM/LL8MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY UNI TIBINRT S5-6RM/LL8MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71422265
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 10/24/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a tka surgery that was performed on (b)(6) 2023, one (1) journey uni tibinrt s5-6rm/ll8mm had dislocated.Patient was brought back to theatre for a liner exchange which was completed successfully.Patient returned to theatre again with another poly dislocation so was revised to tka on (b)(6) 2023.Health status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Additional Manufacturer Narrative
Section h10: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, no relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined based on the limited information provided.Should any additional clinical information be provided, this complaint will be re-evaluated.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed in possible adverse effects that dislocation and subluxation can result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, patient anatomy, postoperative care or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Note: the first revision was reported under your reference number: 1020279-2023-02228 (our reference number: (b)(4)).Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY UNI TIBINRT S5-6RM/LL8MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18171003
Report Number1020279-2023-02272
Device Sequence Number20762539
Product Code HSX
UDI-Device Identifier03596010582249
UDI-Public03596010582249
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K061011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71422265
Device Lot Number19EM09969
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer12/06/2023
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date12/11/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-