U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRAFLEX ESOPHAGEAL NG; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRAFLEX ESOPHAGEAL NG; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00514200
Medical Device Problem Codes Failure to Advance (2524); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block e1: the initial reporter's phone number is (b)(6).Block h6: imdrf device code a15 captures the reportable event of ultraflex esophageal stent partially deployed.
 
Event or Problem Description
It was reported to boston scientific corporation that an ultraflex esophageal covered distal stent was used to treat a 3cm malignant esophageal stenosis during a stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was not dilated prior to stent placement.During the procedure, the device was unable to cross the lesion due to severe stenosis and the stent could not be released when deployment was attempted.The procedure was completed with a different device.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX ESOPHAGEAL NG
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18173089
Report Number3005099803-2023-06301
Device Sequence Number7687632
Product Code ESW
UDI-Device Identifier08714729121190
UDI-Public08714729121190
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091816
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/04/2023
Device Model NumberM00514200
Device Catalogue Number1420
Device Lot Number0027435215
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2023
Initial Report FDA Received Date11/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient Weight65 KG
-
-