| Catalog Number |
680H |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the customer received a thread pack with red marked and opened threads.The packaging blister has the red marking and these blisters are also open.There was no patient involvement.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned a review of the batch manufacturing records was conducted, and no related non-conformances were identified.Related events captured via 2210968-2023-09057, 2210968-2023-09058, 2210968-2023-09059, 2210968-2023-09061, 2210968-2023-09062.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 1/31/2024 component code: g07002 pending evaluation of returned device a device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.Additional information was requested, the following was obtained: the customer reported that there were six samples with red markings.However, upon return of the product, only three samples with the event reported were received.Then we only have three samles - will the other three samples be sent for evaluation? no this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.Part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary: visual inspection was conducted one unopened sample that pertain to product code 680h, lot tabbru.Additionally, a photo was provided and it showed the condition of the reported event.The ethicon san lorenzo team performed a further analysis to determine if the complaint represent a defect/or a process deviation.According to the results: the splice (red tape) can be seen between the copolymer material and the tyvek material by the packaging process at san lorenzo.Based on the information currently available, it was confirmed an open seal defect on the portion where the splice material is present.Therefore, pia escalation was performed to rmb under pie 2315297/pia 2328981 and nr-0205929.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.
|
| |
|