• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE INC. I-STAT ACT CELITE CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE INC. I-STAT ACT CELITE CARTRIDGE Back to Search Results
Catalog Number 03P86-25
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 06-nov-2023, abbott point of care was contacted by a customer regarding act celite cartridges that yielded descrepant results on a 57 year old male patient seen for coronary bypass grafting x2 using left mammary and left endoscopic.There was no additional patient information available at the time of this report.Return product is available for investigation.Date collected tested result (sec) 06-nov-2023 08:44 08:44 139 baseline 06-nov-2023 09:45 09:45 782 06-nov-2023 10:05 10:05 741 06-nov-2023 10:12 10:12 139 heparin times/dose, reversal information provided.At this time there is no reason to suspect a malfunction exists.No repoprts of delay or impact to patient management.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.An investigation is underway.Per i-stat system manual (art: 714185-00q), the intended use of the i-stat celite activated clotting time (celite act) test is an in vitro diagnostic test that uses fresh, whole blood, and is useful for monitoring patients receiving heparin for treatment of pulmonary embolism or venous thrombosis, and for monitoring anticoagulation therapy in patients undergoing medical procedures such as catheterization, cardiac surgery, surgery, organ transplant, and dialysis.
 
Additional Manufacturer Narrative
Apoc incident # 937531.Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Additional Manufacturer Narrative
Apoc incident: (b)(4).The investigation was completed on 12-dec-2023.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained and returned cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Al (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT ACT CELITE CARTRIDGE
Common Device Name
ACT CELITE CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE INC.
400 college road
princeton NJ 08540
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
185 corkstown road
ottawa, ontario K2H 8-V4
CA   K2H 8V4
MDR Report Key18175786
Report Number2245578-2023-00166
Device Sequence Number15382743
Product Code JBP
UDI-Device Identifier10054749000149
UDI-Public10054749000149
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K992571
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2024
Device Catalogue Number03P86-25
Device Lot NumberR23240
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/06/2023
Supplement Date Received by Manufacturer12/12/2023
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
-
-