| Catalog Number |
04404483190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).Calibration was last performed on 23-aug-2023.The qc recovery data provided was acceptable.The customer stated they replaced the glucose reagent and performed qc and a precision check successfully.The service maintenance actions performed by the customer resolved the issue.No further issues were reported.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable gluc3 glucose hk gen.3 results for 3 patient samples on a cobas 6000 c (501) module.The customer was prompted the repeat the samples as the initial results seemed too low.For sample 1, the initial glucose result was 3 mg/dl.The sample was repeated on another c501 analyzer and the result was 215 mg/dl.For sample 2, the initial glucose result was 3 mg/dl.The sample was repeated on another c501 analyzer and the result was 298 mg/dl.For sample 3, the initial glucose result was 2 mg/dl.The sample was repeated on another c501 analyzer and the result was 115 mg/dl.The repeat results were deemed correct.No questionable results were reported outside of the laboratory.
|
| |
|