The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the insertion flexible tube fluid damage, the control body fluid damage, the lcb (light carrying bundle) fluid damage, the lg connector deformed, the lg connector leak, the operation channel (primary) cut, the control body disinfections damage, the forward body cover disinfections damage, the body cover grip disinfections damage, the light guide cable disinfections damage, and the lg cable connector housing disinfections damage; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|