| Model Number |
25 DEG TIGHT CURVE LEFT, QUICKPASS LASSO |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was report that during an arthroscopic suturing of the meniscus, the device blocked and the surgeon was not able to pull the suture.There was no harm for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint is not confirmed.Visual inspection did not find any issues with the device returned.Functional testing with suture 2-0 did go through the shaft to the tip without an issue of the 25°, tight curve, left, quickpass lasso.No problem found.
|
| |
|