| Model Number |
LZ226822320S20 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
Provider alleges the consumer allegedly smelled smoke while sitting in the radiance, but couldn't identify where it was coming from - the consumer allegedly found the chair on fire shortly after this, dragged it outside where it allegedly became engulfed in flames.
|
| |
|
Event or Problem Description
|
|
Provider alleges the consumer allegedly smelled smoke while sitting in the radiance, but couldn't identify where it was coming from - the consumer allegedly found the chair on fire shortly after this, dragged it outside where it allegedly became engulfed in flames.
|
| |
|
Additional Manufacturer Narrative
|
|
Remnant harnessing was examined and no evidence of arcing was found.With this, the origin of the event is unknown.
|
| |
|
Search Alerts/Recalls
|