| Catalog Number |
10736404 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a discrepant low creatinine result compared to retesting of the same? sample on their laboratory instrument.Customer stated that the level 1 qc failed low for creatinine on (b)(6).There is no report of injury due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens requested instrument log files for further investigation.The customer stated that level 1 qc for creatinine failed low the day prior to the event.Siemens does not recommend running patient samples if quality controls are not passing.The cause of this event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The information provided by the customer was reviewed.Without the requested instrument file, a full investigation was unable to proceed.However, a review of the lot performance of the card lot in question at the time of the event was performed.The in-house performance of creatinine for the card lot in question did not identify any product deficiencies.Review of testing performed at release, throughout product lifetime, and additional testing did not show any anomalies that aligns with the observation in the customer allegation.No product problem identified.The cause of this event is unknown.
|
| |
|