| Model Number |
N/A |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer obtained discordant enzymatic creatinine_3 (ecre3) for a patient serum sample processed on the atellica ch 930 analyzer, serial number (s/n) (b)(6) and reagent lot 123531.The discordant result was not reported to the physician.The discordant result was discovered when the customer performed post-analysis/review of results from patient samples with initial low ecre3 results.The customer retested the samples on a different analyzer and sample id 1739795101 was found to be discordant.There are no allegations of patient or user intervention or adverse health consequences due to the discordant ecre3 result.
|
| |
|
Additional Manufacturer Narrative
|
|
An outside of the united states (ous) customer contacted the siemens customer care center to report an observation of a falsely depressed atellica ch enzymatic creatinine3 (ecre3) patient result.Quality control (qc) recovered within acceptable ranges for all control levels.The calibration used in sample processing was acceptable.The interpretation of results section of the instructions for use (ifu) states, "results of this assay should always be interpreted in conjunction with the patient¿s medical history, clinical presentation, and other findings." the limitations section of the ifu states, "as with any chemical reaction, you must be alert to the possible effect of unknown interferences from medications or endogenous substances.The laboratory and physician must evaluate all patient results in light of the total clinical status of the patient." siemens is investigating.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed mdr 1219913-2023-000295 initial report on (b)(6) 2023.Additional information - siemens investigated an outside of the united states (ous) customer observation of a discordant enzymatic creatinine_3 (ecre3) lot 123531 result on atellica ch 930 instrument, serial number (b)(6).Quality control (qc) results were within the published acceptable ranges before and after the discordant result was generated, indicating there is no reagent issue.Siemens evaluated the atellica ch 930 system data and noted erratic kinetic errors for the samples in question.The errors were associated to system autocheck failures indicating a contaminated water bath.Customer did not perform maintenance or troubleshooting actions recommended to resolve this issue at that time.Instead, the customer discontinued patient sample testing.Subsequently, the customer informed siemens that they initiated patient sample testing on (b)(6) 2023.Siemens reviewed the system data and noted that autocheck testing was successful on (b)(6) 2023 indicating that the water bath contamination issue was resolved, and no kinetic errors were noted.Customer is fully operational.In section h6 of this report, the investigation finding and conclusion codes were updated based on the investigation results.
|
| |
|
Search Alerts/Recalls
|