Model Number N/A
Device Problem Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10: 51-104110 item name tprlc 133 t1 pps ho 11x142mm lot # 6883789, 51-103160 item name tprlc 133 t1 pps so 16x152mm lot# 6508849.G2: foreign: japan.Multiple mdr reports were filed for this event, please see associated reports 0001825034 - 2023 - 02713, 0001825034 - 2023 - 02714.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : device in process to return.
 
Event Description
It was reported that when inspecting item in stock, it was found that the packaging was damaged.There is no additional information available at the time of this report.
 
Manufacturer Narrative
Upon reassessment of the reported event, it was determined to be not reportable.The initial report was forwarded in error.Updated: b4, b5, g3, g6, h2, h6, h10.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been breached.The reported event has been confirmed by evaluation of the returned product and provided photos.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
Upon reassessment of the reported event, it was determined to be not reportable.The initial report was forwarded in error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 14X148MM
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18181762
MDR Text Key328661002
Report Number0001825034-2023-02712
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00880304517158
UDI-Public(01)00880304517158(17)291022(10)6656596
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-105140
Device Lot Number6656596
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/31/2023
Initial Date FDA Received11/21/2023
Supplement Dates Manufacturer Received02/01/2024
Supplement Dates FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
PLEASE SEE H10
Patient SexPrefer Not To Disclose
-
-