• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC CR TIB INSERT SLOPE +. SZ 3.5, 9 MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC CR TIB INSERT SLOPE +. SZ 3.5, 9 MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-48-3509
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 11/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 4 years post op initial left tka, this 73 y/o female patient was revised due to poly wear and femur loosening.Patient was last known to be in stable condition following the event.Product not returning - disposed.
 
Additional Manufacturer Narrative
Section h10: (h3) pending evaluation; (d10) concomitant device(s): 5777031 - 02-010-04-0235 - logic cr femoral por, left, sz 3.5; 3933664 - 02-012-45-3525 - lgc tibial fit tray cem sz 3.5f / 2.5t; 5939545 - 200-02-35 - three peg patella 35mm; 5719606 - 201-78-15 - holding pin mini sharp point 4 pk; 5776148 - 201-78-81 - 3 trocar, mod.Hex 2pk; s019825 - 521-78-23 - threaded pin size 2.3 collared; s032419 521-78-23 - threaded pin size 2.3 collared; s031429 521-78-32 - threaded pin size 3.0 collarless; 6028019001 a10012 - gps implant kit v2.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of presumed prosthesis wear of the tibial insert.However, images of the revised insert do not show signs of excessive prosthesis wear inconsistent with being implanted.The aseptic (non-infected) femoral loosening was likely the result of an insufficient bond between the femoral component and the bone.However, this cannot be confirmed because the revised components were not returned for evaluation and no radiographs were provided.
 
Additional Manufacturer Narrative
*the following sections have corrected information: (h6) type of investigation, investigation findings, investigation conclusions.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported was likely the result of presumed prosthesis wear of the tibial insert.However, images of the revised insert do not show signs of excessive prosthesis wear inconsistent with being implanted.The aseptic (non-infected) femoral loosening was likely the result of an insufficient bond between the femoral component and the bone.However, this cannot be confirmed because the revised components were not returned for evaluation and no radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC CR TIB INSERT SLOPE +. SZ 3.5, 9 MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 ct
gainesville, FL 32653
3523771140
MDR Report Key18181908
Report Number1038671-2023-02831
Device Sequence Number7661640
Product Code JWH
UDI-Device Identifier10885862159281
UDI-Public10885862159281
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111400
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/17/2022
Device Catalogue Number02-012-48-3509
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2023
Supplement Date Received by Manufacturer02/27/2024
10/07/2024
05/05/2025
Initial Report FDA Received Date11/21/2023
Supplement Report FDA Received Date03/04/2024
10/09/2024
05/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/19/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0021-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient Age73 YR
Patient SexFemale
-
-