OPSENS SAVVYWIRE; PRE-SHAPED PRESSURE GUIDEWIRE Back to Search Results
Model Number F3001
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Cardiac Tamponade (2226); Pericardial Effusion (3271)
Date of Event 10/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation of the wire confirmed the description of the event, a fracture occurred in the wire's tip.The tip also presented a prominent kink 6 mm from the distal end of the wire, the kink occurred while snaring the remain of the tip from the left ventricle.Visual inspection of the wire confirmed that the structural assembly of the wire was done within specification.Check of the device history record confirmed that savvywire lot number osw-0019 was released per specifications.No non-conformity nor deviation were associated with this lot.A possible damage during unpacking, or while advancing the wire could not be determined during this investigation.The following precaution are mentioned in the ifu : never advance, pull or torque a savvywire against resistance without first determining the reason for the resistance under fluoroscopy.Excessive force against resistance may result in damage to the wire and/or to the anatomy.Fluoroscopic guidance should be used during manipulations of the savvywire.Care should be taken when the tip of the savvywire is positioned, moved or torqued in the ventricle.Unfortunately, the hospital couldn't provide the case angio images from before the break of wire following opsens's request.Without the angio images, opsens is not able to fully investigate and conclude the exact timing and circumstances of the fracture, therefore no definitive root cause can be assigned for the event.If there is any further relevant information provided, a follow up report will be submitted.
 
Event or Problem Description
Opsens was informed on the (b)(6) that savvywire was introduced into the lv through the 6f pigtail, as the physician was advancing the z-med balloon over the savvywire, it was noticed under the fluro that the wire was broken at the tip.After several attempts the physician was able to snare the wire out of patient's body.Tte revealed that the patient had an effusion and tamponade.Subsequently, blood transfusion was done, and patient status was stabilised.Opsens later reached out to the physician for additional information regarding the case and learned that the patient expired a couple of days after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAVVYWIRE
Common Device Name
PRE-SHAPED PRESSURE GUIDEWIRE
Manufacturer (Section D)
OPSENS
750
blvd du parc technologique
quebec, quebec G1P 4 S3
CA  G1P 4S3
Manufacturer (Section G)
OPSENS
750
blvd du parc technologique
quebec, quebec G1P 4 S3
CA   G1P 4S3
Manufacturer Contact
maryem oudhini
750
blvd du parc technologique
quebec, quebec G1P 4-S3
CA   G1P 4S3
MDR Report Key18181924
Report Number3008061490-2023-00012
Device Sequence Number18064111
Product Code DXO
UDI-Device Identifier07540184030019
UDI-Public(01)07540184030019(11)240628(17)220628(10)OSW-0019
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K213854
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2024
Device Model NumberF3001
Device Lot NumberOSW-0019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2023
Initial Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-