• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Medical Device Problem Codes Fracture (1260); Device Damaged by Another Device (2915); Activation Failure (3270)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that stent fracture and damage post-deployment occurred requiring intervention.The patient presented with coronary artery disease (cad).Percutaneous coronary intervention (pci) was performed.Following pre-dilation, a 4.00 x 16mm synergy xd drug-eluting stent was deployed in ostial right coronary artery.The stent delivery system was then removed, and angioplasty was performed.During the angioplasty, it was observed that the proximal stent was in question of expansion, and it was decided to perform an intravascular ultrasound (ivus) using non-boston scientific (bsc) device.Under ivus, it was noted that the stent had not fully expanded.A 4.5x12 nc quantum apex was then placed and would not expand the stent either, so the physician decided to use a shockwave with a 4.0x12 sw balloon.Upon inflation and delivery of the shocks, it was noted that the stent had a malformation.The shockwave balloon was removed, and the proximal edge part of the stent with the crown was snapped in half.The stent remained in the patient and was scheduled for open heart surgery to retrieve the broken part of the stent.The patient was stable post-procedure and is expected to fully recover.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18182206
Report Number2124215-2023-65139
Device Sequence Number20805665
Product Code NIQ
UDI-Device Identifier08714729981121
UDI-Public08714729981121
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2023
Initial Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-