• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3® Ø12X170MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3® Ø12X170MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 31252170S
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "intraoperative interference between the step drill and the nail was observed.The nail was replaced and the procedure was completed without problems.".
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was returned, but found to be functional.The received nail was visually inspected and found to be in good condition overall.A functional inspection was conducted with the returned nail, target device, speed lock sleeve, and nail holding screw, with a sample drill and sample drill sleeve, in which we can see if the speed lock is locked at the proper angle, which matches with the nail there is no misdrilling.If there is a mismatch between the angle of the nail and the angle locked at the target device and speed lock, there will be a mismatch.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the above investigation the root cause can be attributed to user related as a functional inspection was performed with a returned nail that proves that the nail is perfectly functional.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
As reported: "intraoperative interference between the step drill and the nail was observed.The nail was replaced and the procedure was completed without problems.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø12X170MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18182537
Report Number0009610622-2023-00428
Device Sequence Number15402021
Product Code HSB
UDI-Device Identifier04546540375094
UDI-Public04546540375094
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number31252170S
Device Lot NumberK05F94F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/27/2023
Supplement Date Received by Manufacturer12/14/2023
Initial Report FDA Received Date11/21/2023
Supplement Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-