| Catalog Number |
2872000012 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Pressure Sore/Ulcer (2326)
|
| Date of Event |
11/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, while using the surface, the patient complained of pain in their sacral area.It was discovered that the surfaces gel had separated, and the patient was sitting on the frame of the mattress.The patient sustained a new stage 2 pressure injury.Attempts are being made to gather additional details from the user facility.
|
| |
|
Event or Problem Description
|
|
It was reported that, while using the surface, the patient complained of pain in their sacral area.It was discovered that the surfaces gel had separated, and the patient was sitting on the frame of the mattress.The patient sustained a new stage 2 pressure injury.
|
| |
|
Additional Manufacturer Narrative
|
|
This event was previously evaluated through an internal investigation of the failure mode and found to not pose any additional risk to the patient.From this information, a stryker quality assurance engineer determined that the stage 2 pressure injury resulting from uneven pressure distribution was likely not due to any component level malfunction as the gel issue alleged.This issue was resolved for the customer by replacing the surface.
|
| |
|
Search Alerts/Recalls
|