| Model Number |
RNS 320 |
| Medical Device Problem Codes |
Failure to Interrogate (1332); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient had her device placed in mri mode and then underwent an mri.Following the mri, the neurostimulator would not communicate with a programmer nor the remote monitor on (b)(6) 2023 the neurostimulator was replaced and returned to neuropace for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The explanted device was returned to neuropace and is undergoing analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) investigation this device experienced an externally impressed voltage during an mri procedure.The cinco 1e asic in this device went to a non-functional state wherein the internal regulated power supplies are stable at nearly zero volts.Cycling the power returned the device to normal operation.This is a known behavior of the cinco 1e asic.
|
| |
|
Event or Problem Description
|
|
Original report- the patient had her device placed in mri mode and then underwent an mri.Following the mri, the neurostimulator would not communicate with a programmer nor the remote monitor.On (b)(6) 2023 the neurostimulator was replaced and returned to neuropace for investigation.
|
| |
|