SYNTHES GMBH RFNA / 14MM / 380MM STANDARD BEND / STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.233.438S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Nonunion/Delayed-union Bone Fracture (2369); Limb Fracture (4518)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: hwc.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a photo review.Review of the x-ray evidence found nothing indicative of a device nonconformance or a relation with the reported adverse event.Through follow up it is now understood there is no allegation associated against a product malfunction.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): manufacturing location: monument.Manufacturing date: 30-mar-2022.Expiration date: unknown.Part number: 04.233.438s, rfn/14mm/380mm 5° bend/pe inlay/sterile.Lot number: 579p490 (sterile).Lot quantity:(b)(4).One piece was scrapped in cell for a faded/incomplete etch.Production order traveler met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, in-process / inspect dimensional / final met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, rfn assembly inspection met all inspection acceptance criteria.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn was recorded on the production order traveler.The scn supplied by fruh was reviewed and met specified dose ranges.There was no pll included in this dhr.Therefore, the pll for this lot could not be reviewed and the expiration date could not be notated.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to non-union and femoral neck fracture¿ does not indicate breakage of the nail or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(4) 2023, there was a removal of an rfna (retrograde femoral nail advanced) due to non-union and a femoral neck fracture.The patient underwent a total hip arthroplasty (tha) and revision nail implantation.This report involves one rfna / 14mm / 380mm standard bend / sterile.This is report 1 of 7 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFNA / 14MM / 380MM STANDARD BEND / STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18184372
Report Number8030965-2023-14651
Device Sequence Number7611503
Product Code HSB
UDI-Device Identifier10886982291165
UDI-Public(01)10886982291165
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K201346
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number04.233.438S
Device Lot Number579P490
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2023
Initial Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: 4.5 MM; UNK - SCREWS: 5.0 MM LOCKING; UNK - SCREWS: 5.0 MM LOCKING; UNK - SCREWS: 5.0 MM LOCKING; UNK - SCREWS: 5.0 MM LOCKING; UNK - SCREWS: 5.0 MM LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-