Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: hwc.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a photo review.Review of the x-ray evidence found nothing indicative of a device nonconformance or a relation with the reported adverse event.Through follow up it is now understood there is no allegation associated against a product malfunction.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): manufacturing location: monument.Manufacturing date: 30-mar-2022.Expiration date: unknown.Part number: 04.233.438s, rfn/14mm/380mm 5° bend/pe inlay/sterile.Lot number: 579p490 (sterile).Lot quantity:(b)(4).One piece was scrapped in cell for a faded/incomplete etch.Production order traveler met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, in-process / inspect dimensional / final met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, rfn assembly inspection met all inspection acceptance criteria.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn was recorded on the production order traveler.The scn supplied by fruh was reviewed and met specified dose ranges.There was no pll included in this dhr.Therefore, the pll for this lot could not be reviewed and the expiration date could not be notated.This lot met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to non-union and femoral neck fracture¿ does not indicate breakage of the nail or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|