| Model Number |
0600-0767 |
| Medical Device Problem Codes |
Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer declined the option to have the reported glidescope core quickconnect cable returned to verathon for evaluation after isolating the issue to the cable.Since the device was not returned to verathon for evaluation, the cause could not be determined.A complaint history search could not be performed for the reported glidescope core quickconnect cable due to the lot number not being provided to verathon.Trending analysis for the glidescope core quickconnect cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized by verathon.The customer reported replacing the subject glidescope core quickconnect cable following the reported event.Corrective action is currently not required at this time.Verathon will continue to monitor for any ongoing trends.
|
| |
|
Event or Problem Description
|
|
A customer reported that during an emergency care procedure, using a glidescope core quickconnect cable, black and grey scrambled lines appeared on the glidescope core 15-inch monitor when a glidescope bflex 5.8 single-use bronchoscope was connected.A different glidescope bflex 5.8 single-use bronchoscope was connected to the cable and monitor but the black and grey scrambled lines appeared again repeatedly after the bronchoscope was placed into the patient's airway.The procedure was completed using a different glidescope core system with the same bronchoscope used on the patient previously.No delay in the procedure or harm to the patient was reported.Follow-up information received from the customer reported that they were able to isolate the issue to the glidescope core quickconnect cable that was used with the initial monitor and bronchoscope.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a supplemental report correcting the unique device identifier (udi) information in section d4.In the initial report submission, the udi information field was left blank due to the subject device lot number and manufacture date being unknown.Despite verathon being unable to obtain the lot information from the customer, this supplemental report provides the udi-di for the subject device model.
|
| |
|