| Model Number |
SUTR ANCHBIO-COMP S-TAK KNOTLESS |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 10/26/2023, it was reported by a sales representative via email that (3) ar-3636 knotless fibertak and an ar-1938bc knotless biocomposite suturetak pulled out of implantation site.The surgeon drilled new bone socket each time the anchors pulled out.The case was completed successfully using a pushlock.This was discovered during a bankart repair procedure.The case was delayed about twenty minutes.Additional information received on 10/26/2023: the procedure took place on 10/25/2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely causes of the reported failure are improper bone preparation, misaligned insertion, and prying/leveraging the device during insertion.
|
| |
|