| Lot Number |
0000320231 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Codes |
Obstruction/Occlusion (2422); Embolism/Embolus (4438)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A2: age at time of event: over 18 years.
|
| |
|
Event or Problem Description
|
|
It was reported that obsidio embolized.The target location was in the gastroduodenal artery (gda).The vessel size was 1 - 3mm.Obsidio was selected for use.After creating casting of the obsidio, the fellow was instructed by the physician to administer a large bolus of saline and obsidio material through aliquot technique.The reason for the use of the aliquot method was to re-use the microcatheter.The aliquot method resulted in a total of.6 ml of obsidio being used and the catheter, in use, being placed too close to the obsidio casting.The obsidio from the second bolus/aliquot, moved towards the branch point of the right hepatic artery.The flow of the vessel and injection of contrast during a hand injection run took the obsidio even more distal.The obsidio shut down the right hepatic artery supply.Action was taken to break up the obsidio into pieces using a microcatheter.Smaller pieces of obsidio traveled distally to the pancreas.Minor flow was achieved but not entirely.Imaging, post procedure, revealed the patient had pancreas flow shut down resulting into other complications like pancreas infarct.
|
| |
|
Additional Manufacturer Narrative
|
|
A2: age at time of event: over 18 years.Image evaluation by mfr: the complaint device was not received for analysis.The site provided images of the embolized vessel, confirming the events of vessel occlusion and embolism.H3 other text: no device was returned.Images were reviewed.
|
| |
|
Event or Problem Description
|
|
It was reported that obsidio embolized.The target location was in the gastroduodenal artery (gda).The vessel size was 1 - 3mm.Obsidio was selected for use.After creating casting of the obsidio, the fellow was instructed by the physician to administer a large bolus of saline and obsidio material through aliquot technique.The reason for the use of the aliquot method was to re-use the microcatheter.The aliquot method resulted in a total of.6 ml of obsidio being used and the catheter, in use, being placed too close to the obsidio casting.The obsidio from the second bolus/aliquot, moved towards the branch point of the right hepatic artery.The flow of the vessel and injection of contrast during a hand injection run took the obsidio even more distal.The obsidio shut down the right hepatic artery supply.Action was taken to break up the obsidio into pieces using a microcatheter.Smaller pieces of obsidio traveled distally to the pancreas.Minor flow was achieved but not entirely.Imaging, post procedure, revealed the patient had pancreas flow shut down resulting into other complications like pancreas infarct.
|
| |
|
Event or Problem Description
|
|
It was reported that obsidio embolized.The target location was in the gastroduodenal artery (gda).The vessel size was 1 - 3mm.Obsidio was selected for use.After creating casting of the obsidio, the fellow was instructed by the physician to administer a large bolus of saline and obsidio material through aliquot technique.The reason for the use of the aliquot method was to re-use the microcatheter.The aliquot method resulted in a total of.6 ml of obsidio being used and the catheter, in use, being placed too close to the obsidio casting.The obsidio from the second bolus/aliquot, moved towards the branch point of the right hepatic artery.The flow of the vessel and injection of contrast during a hand injection run took the obsidio even more distal.The obsidio shut down the right hepatic artery supply.Action was taken to break up the obsidio into pieces using a microcatheter.Smaller pieces of obsidio traveled distally to the pancreas.Minor flow was achieved but not entirely.Imaging, post procedure, revealed the patient had pancreas flow shut down resulting into other complications like pancreas infarct.It was further reported that the gda was treated additionally with two 4mmx6cm coils.Subsequent contrast injection demonstrated no further extravasation and no flow to the gda.Diminished hepatic artery (rha) flow due to non-target embolization was also noted.Attempts were made to macerate the migrated conformable embolic with vigorous saline flushes via the microcatheter.Balloon maceration was also performed with a balloon occlusion microcatheter.This resulted in improved but persistently diminished flow within the rha and may have led to additional non-target embolization.
|
| |
|
Additional Manufacturer Narrative
|
|
Image evaluation by mfr: the complaint device was not received for analysis.The site provided images of the embolized vessel, confirming the events of vessel occlusion and embolism.
|
| |
|