• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIOSURE HA 8MM X 25MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIOSURE HA 8MM X 25MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 72201776
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an acl reconstruction, the biosure screw cracked.The procedure was completed with a smith and nephew back up device and it is unknown if there was a surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
H10: h2: additional information a1, a2, a3, a4, b5, e1 and h6.
 
Event or Problem Description
It was reported that during an acl reconstruction, the biosure screw cracked.The original drilled bone hole was used and no piece fell into the patient's wound.The procedure was completed with a smith and nephew back up device with no surgical delay.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection revealed that it is not in its original packaging.The proximal end of the screw is deformed and there are broken threads along the length of the device.There is biological debris on the returned item.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specification, found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE HA 8MM X 25MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18184704
Report Number1219602-2023-02260
Device Sequence Number15375301
Product Code HWC
UDI-Device Identifier03596010611727
UDI-Public03596010611727
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K080358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201776
Device Lot Number51005198
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2023
Supplement Date Received by Manufacturer11/20/2023
02/14/2024
Initial Report FDA Received Date11/21/2023
Supplement Report FDA Received Date11/22/2023
02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age21 YR
Patient SexMale
Patient Weight66 KG
-
-