FX SHOULDER SOLUTIONS DALLAS HUMELOCK REVERSED; REVERSE SHOULDER PROSTHESIS Back to Search Results
Catalog Number 317-3614
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 11/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient was revised 1 month after primary surgery due to lack of following surgical technique during primary surgery.No actual, implied, or suspect link to fx product.
 
Event or Problem Description
Revision was performed approximately 1 month after the primary surgery; primary surgery occurred on (b)(6) 2023.No fall or other trauma noted.Patient presented with a loosened stem within the humeral canal.36x14 stem, 40 mm +3 stability humeral cup, and 9 mm spacer explanted.These parts were replaced with a 40x14 stem, 40 mm + 3 stability humeral cup, 9 mm spacer, a 34 mm cortical screw, and a 28 mm cortical screw.The primary surgery was treatment for a fracture.Contrary to the surgical technique for the primary surgery, the surgeon did not include any distal screws, which allowed the stem to loosen.Surgeon stated that all future proximal humeral fractures will be treated with distal screws.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMELOCK REVERSED
Common Device Name
REVERSE SHOULDER PROSTHESIS
Manufacturer (Section D)
FX SHOULDER SOLUTIONS DALLAS
13465 midway rd
suite 100 & 101
dallas TX 75244
Manufacturer (Section G)
FX SHOULDER SOLUTIONS DALLAS
13465 midway rd
suite 100 & 101
dallas TX 75244
Manufacturer Contact
tristan mahler
13465 midway rd
suite 100 & 101
dallas, TX 75244
7137326920
MDR Report Key18187434
Report Number3014128390-2023-00042
Device Sequence Number6663000
Product Code PHX
UDI-Device Identifier03701037303757
UDI-Public03701037303757
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K162455
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number317-3614
Device Lot NumberS0698
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PART: 103-0007 LOT: P1280; PART: 104-1003 LOT: P0821
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
-
-