MAUDE Adverse Event Report: BECTON DICKINSON AND CO. BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND CO. BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE Back to Search Results
Catalog Number 324910
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using the bd veo¿ insulin syringes with bd ultra-fine¿ needle a black spot is observed inside the syringe hub.The following information was provided by the initial reporter: 1.Before loading the drug into the syringe, a white fm was found inside the barrel.2.A black spot is observed inside the syringe hub.
 
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes d.9.Returned to manufacturer on: 17nov2023.H.6.Investigation summary: samples were received and an investigation was performed.This is the 6th complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Event or Problem Description
It was reported that while using the bd veo¿ insulin syringes with bd ultra-fine¿ needle a black spot is observed inside the syringe hub.The following information was provided by the initial reporter: 1.Before loading the drug into the syringe, a white fm was found inside the barrel.2.A black spot is observed inside the syringe hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VEO¿ INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key18188176
Report Number9616656-2023-01200
Device Sequence Number7275030
Product Code FMF
UDI-Device Identifier00382903249107
UDI-Public00382903249107
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K024112
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2025
Device Catalogue Number324910
Device Lot Number0189393
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2023
Supplement Date Received by Manufacturer03/28/2024
Initial Report FDA Received Date11/21/2023
Supplement Report FDA Received Date04/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-