|
This device is classified as import for export, therefore 510k is not applicable.Model ec34-i10tl-us is available in the usa with a 510k number k180285.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the objective lens cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the objective lens.In addition, our technician confirmed that the angle wire hard to move; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|