| Model Number |
EPK-I7010 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the image noise.Based on the result, we concluded that it was caused due to the excessive shock applied on the video processer.In addition, our technician confirmed that the scope connector mechanical unit loose; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated not to submit mdr.
|
| |
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.Video image failure.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information.B4: date of this report.G6: follow up #1.H2:if follow-up, what type? there were typos in some of the investigation results, which have been corrected.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the image noise.Based on the result, we concluded that it was caused due to the excessive shock applied on the video processer.In addition, our technician confirmed that the scope connector mechanical unit loose; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|