| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: canada.Visual review of the returned broach confirms lot etch.Post is confirmed to be fractured from the broach.All pieces were returned.Cutting edges of teeth show wear from previous uses.Proximal teeth show deformation damage.No other damage was noted.A review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the broach piece dropped off while the surgeon was using the calcar planer.There were no known consequences to the patient.Attempts have been made and no further information has been provided.
|
| |
|