Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G2: foreign: canada.Visual review of the returned broach confirms lot etch.Post is confirmed to be fractured from the broach.All pieces were returned.Cutting edges of teeth show wear from previous uses.Proximal teeth show deformation damage.No other damage was noted.A review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the broach piece dropped off while the surgeon was using the calcar planer.There were no known consequences to the patient.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVENIR CMPL BROACH SIZE 4
Common Device Name
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18190539
Report Number0001822565-2023-03276
Device Sequence Number20468363
Product Code LZO
UDI-Device Identifier00889024510142
UDI-Public(01)00889024510142(11)181019(10)64907740
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K182048
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number574999040
Device Lot Number64907740
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2023
Initial Report FDA Received Date11/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-