• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL COMPONENT PRECOAT RIGHT MEDIAL/LEFT LATERAL SIZE 3; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL COMPONENT PRECOAT RIGHT MEDIAL/LEFT LATERAL SIZE 3; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Arthritis (1723); Metal Related Pathology (4530)
Date of Event 11/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient was revised due to metallosis.On revision, the medial compartment was felt to be over-stuffed and with inadequate tibial slope.The patient had been dropping off the back of the tibial component and eroded the poly.Some metallosis was discovered.The implants were otherwise well fixed.There is no additional information available.
 
Additional Manufacturer Narrative
(b)(4), g2- australia h3- customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the provided pictures identified product that had been implanted.The articular surface looks worn, and the tibial plate and femoral implant have biological material on them.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: fit and alignment appropriate, no evidence of metallosis, wear on bearing.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.This complaint for implant wear is confirmed via provided photos, however metallosis could not be confirmed.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL COMPONENT PRECOAT RIGHT MEDIAL/LEFT LATERAL SIZE 3
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18190808
Report Number0001822565-2023-03235
Device Sequence Number15114984
Product Code HSX
UDI-Device Identifier00889024193635
UDI-Public(01)00889024193635(17)270831(10)63754112
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K033363
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00584200302
Device Lot Number63754112
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2023
Supplement Date Received by Manufacturer04/04/2024
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date04/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight80 KG
-
-