Di: (b)(4).Hologic is submitting this report in accordance with 21.Cfr part 803.The submission is based upon the currently available information.The submission of this report does not represent a confirmation of device malfunction involving the hologic products in the event described.D4: lot and serial number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
|
It was reported that on (b)(6) 2021 a biozorb was implanted, and the patient reported suffering from a hard, painful lump near the site of the biozorb device.The patient also reported suffering from swelling, discomfort, pain, itching, and a burning sensation at and around the site of the biozorb device.The patient also reported believing that the biozorb device has failed to properly absorb.No additional information available.
|
|
After additional information review, it was reported that on (b)(6) 2021, an approximately 48-year-old female with unknown past medical history underwent left partial mastectomy with biozorb placement and left sentinel node biopsy.Around (b)(6) 2023 the patient had a breast ultrasound on the left for "evaluation of a palpable abnormality in the left breast along the lumpectomy scar." this revealed, "left breast lumpectomy bed at site of palpable concern" and recommended clinical follow up.There are no additional details in the medical records available.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
|
|
An investigation was conducted: brief description: it was reported that on (b)(6) 2021 a biozorb was implanted, and the patient reported suffering from a hard, painful lump near the site of the biozorb device.The patient also reported suffering from swelling, discomfort, pain, itching, and a burning sensation at and around the site of the biozorb device.The patient also reported believing that the biozorb device has failed to properly absorb.As a result of the pain and complications of the biozorb device, the patient fears the possibility of another tumor every day, causing significant emotional distress.Investigation methods and findings: the sample was not returned to the post market quality laboratory for further investigation; therefore, the inspection according to the biozorb post market instructions was not able to be conducted for this complaint.Additionally, there was limited patient-related information provided for this event; consequently, a comprehensive investigation could not be conducted.The harms identified, based on clinical symptoms and hazardous situations per for this complaint are: pain, serious (pain, sleep dysfunction, device palpability, failure to resorb).Hypersensitivity/allergic reaction, serious (redness/erythema, itching sensation).Inflammation, serious (limited mobility, lymphedema, swelling).Cause/root cause: the described issue was investigated through a root cause analysis conducted under capa.The initial purpose of this capa was to evaluate the biozorb product for a root cause linking the product design to the reported complaint issues.The results of the root cause assessment did not find any direct link between the device and the reported complaint issues.Consequently, the implementation activities from capa were considered no longer applicable since hologic has ceased selling the biozorb product with no intention of returning it to market.In addition, capa is currently evaluating the biozorb product¿s entire dhf and is actively under remediation.Furthermore, a voluntary recall for the biozorb product was initiated on (b)(6) 2024.Conclusion: the reported issue could not be confirmed as no sample was returned for evaluation.The risk trending analysis does not increase the risk zone index.The potential root cause analysis was assessed under capa.Capa is currently evaluating the biozorb product¿s entire dhf and is actively under remediation.The risks associated with this event are further detailed and evaluated against the product and its hazard analysis within the health risk assessment biozorb, complaint evaluation.Future events will be monitored and trended.Complaint confirmed: no.Device history record (dhr) review: the dhr was reviewed for the corresponding lot, and no quality figures were identified.
|