|
The reported event was inconclusive as no sample was returned for evaluation.It is unknown whether the device had met relevant specifications.It was unknown whether the product had caused the reported failure.Potential root cause for this failure mode could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/ operator error).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon burst.Do not aspirate urine through the drainage funnel wall.Single patient use only.Do not reuse and resterilize.For urological use only.Use luer slip syringe.Do not use needle.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen.If permitted by hospital protocol, the may be cut off.If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|