| Catalog Number |
0035640 |
| Medical Device Problem Codes |
Disconnection (1171); Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the pediatric catheter was sliding out too easily from the collection container, compromising sterility.
|
| |
|
Event or Problem Description
|
|
It was reported that the pediatric catheter was sliding out too easily from the collection container, compromising sterility.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "inlet port of the cap / catheter (fr.) thickness out of specification".The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warnings the catheter should never be forced if resistance is felt.Trauma to the bladder (false passage) can be prevented by adequately lubricating the catheter and stretching the penis to straighten the urethra.Sterile: sterile unless package is opened or damaged." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Search Alerts/Recalls
|