• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION MICRO-VOLUME RADIATION STERILIZED EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION MICRO-VOLUME RADIATION STERILIZED EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N3348
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tubing of a non-dehp micro-volume extension set was found in half.This was identified after opening the packaging.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was performed using the naked eye and it was noted that the blister package was open; the sealing of the blister package between the top web and bottom web was not complete.Additionally the tubing of the set was observed to be damaged (cut).The reported condition was verified.The cause of the reported condition was a manufacturing issue; potentially caused by the package sealing machine.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO-VOLUME RADIATION STERILIZED EXTENSION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - AIBONITO
rd 721 km 0 3 po box 1389
aibonito 00705
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18192377
Report Number1416980-2023-06123
Device Sequence Number17953666
Product Code FPA
UDI-Device Identifier00085412004822
UDI-Public(01)00085412004822
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K811078
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2N3348
Device Lot NumberUR23F09161
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2023
Supplement Date Received by Manufacturer12/27/2023
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date12/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-