| Catalog Number |
2N3348 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tubing of a non-dehp micro-volume extension set was found in half.This was identified after opening the packaging.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.A visual inspection was performed using the naked eye and it was noted that the blister package was open; the sealing of the blister package between the top web and bottom web was not complete.Additionally the tubing of the set was observed to be damaged (cut).The reported condition was verified.The cause of the reported condition was a manufacturing issue; potentially caused by the package sealing machine.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|