|
|
| Catalog Number |
AB35W06150135 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician attempted to use an evercross pta balloon with a 6fr non-medtronic sheath and a 45cm non-medtronic guidewire during treatment of a 25cm calcified lesion in the patients mid left superficial femoral artery (sfa).Little vessel tortuosity and moderate vessel calcification are reported.Lesion exhibited 75% stenosis.Artery diameter reported as 6mm.There were no abnormalities reported in relation to anatomy.A spider fx embolic protection device was used.A non-medtronic inflation device was used for balloon inflation.There was no damage noted to packaging, i.E.Shelf carton, hoop/tray.No issues were noted when removing the device from the hoop/tray.Ifu was followed and the device was prepped without issue.The device was not passed through a previously deployed stent.No resistance was noted during advancement of the device.No excessive force was used.A longitudinal balloon burst was reported during balloon inflation at 8atm.Balloon was removed in one piece with all fragments still partially attached to the catheter, all fragments of the balloon were safely removed from the patient.No vessel damage noted.Alternative pta balloon angioplasty was used to complete the procedure.No patient injury reported.
|
| |
|
Search Alerts/Recalls
|
|
|