MAUDE Adverse Event Report: COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNIVERSA FIRM URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Catalog Number UFH-630-R
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the string (tether) on the universa firm ureteral stent broke.There were no adverse effects to the patient reported.Additional information has been requested regarding the patient and the event.At the time of this report, no further information has been provided.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.E3: customer occupation = (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Additional information: b5; h6: health effect - clinical code (annex e) and health effect - impact code (annex f).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported in additional information that the nurse (rn) opened the the device packaging to the sterile field, and the staff noticed that the string was not attached to the stent.The device was then removed from the field and sequestered.The device did not make patient contact, and a new stent was was obtained to complete the procedure.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: d9, h3 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: it was reported that the string (tether) on a universa firm ureteral stent broke.The rn opened the device packaging to the sterile field, and the staff noticed that the string was not attached to the stent.The device was then removed from the field and sequestered.The device did not make patient contact, and a new stent was obtained to complete the procedure.There were no adverse effects from this incident as the device did not make contact with the patient.Reviews of documentation including the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.A device failure analysis was conducted on the returned stent.The stent was returned without the tether.Aside from a kink in the stent due to repackaging, no other damage was noted.The complaint was confirmed as the tether was not attached to the returned device.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot and the related subassembly lot revealed no nonconformances.A complaint history search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The product ifu, [t_uss_rev5] "universa firm ureteral stents and stent sets" does not contain information related to the reported failure mode.Based on the information provided, inspection of the returned device, and the results of the investigation, the cause of the separation was unable to be determined.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSA FIRM URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18192907
Report Number1820334-2023-01610
Device Sequence Number17972244
Product Code FAD
UDI-Device Identifier10827002499086
UDI-Public(01)10827002499086(17)260823(10)15624182
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2026
Device Catalogue NumberUFH-630-R
Device Lot Number15624182
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2023
Supplement Date Received by Manufacturer11/24/2023
01/08/2024
05/28/2024
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date12/15/2023
02/02/2024
06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-