MAUDE Adverse Event Report: MICROVENTION INC TRAXCESS GUIDEWIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION INC TRAXCESS GUIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number GW1420040
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was stated to be available for return to the manufacturer for evaluation but has not yet been returned.The alleged product issue / foreign material as described could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.
 
Event or Problem Description
It was reported that during preparation, once the guidewire was hydrated, they proceeded to insert it into the microcatheter through the guide pin.They noticed it released tiny dark-colored material.The device was not used.Although there was no patient involvement or contact, it is possible that the foreign material/contamination could potentially go undetected and could cause or contribute to a death or serious patient injury should the event recur.Therefore, the event is being reported.
 
Event or Problem Description
Additional information received notes they normally use headway, neuroslider, echelon microcatheters, although it is unknown which was used during this incident.The preparation process was explained as follows: as usual, they hydrated the traxcess inside the container tube and in the tray, and then, they passed the traxcess through the microcatheter.They noticed some kind of friction between the traxcess and the microcatheter that they were not used to.No other clarification was provided.
 
Additional Manufacturer Narrative
B5: additional information provided d10: device received h10: investigation: actual sample received: one guidewire; one mandrel.Visual inspection of the guidewire: there was no foreign material adhering to the guidewire.Magnifying inspection of the guidewire: there was no coil disturbance or kink in the platinum coil section and stainless steel coil section.Intermittent peeling of ptfe coat was observed on the following areas of the shaft: approx.1140 mm -approx.1510mmfrom the distal end; approx.1630 mm - approx.1730mmfrom the distal end; approx.1770 mm - approx.1850mmfrom the distal end; there were no scratches or other anomalies on the other part of shaft section.Dimensions of the guidewire: outer diameter of the platinum coil section: it met the factory's specifications.No anomaly was found.Outer diameter of the stainless steel coil section: it met the factory's specifications.No anomaly was found.Outer diameter of the shaft section: it met the factory's specifications.No anomaly was found.Visual inspection of the mandrel: there was foreign substance adhering to the distal end of mandrel.From the description of the event, this was thought to be ¿tiny dark-colored material¿ released from the guidewire.Magnifying inspection of the foreign substance: it was a blue thread-like substance.Component analysis of the foreign substance (ft-ir*) the peak of the foreign substance was similar to the peak of the ptfe coat.Therefore, it was considered that the tiny dark-colored material pointed out in this complaint was a piece of ptfe coat peeled off from the actual sample.*ft-ir (fourier transform infrared spectroscopy method): an analysis method to specify the molecular structure of an object by analyzing the spectrum obtained by irradiating the target with infrared rays.Since the wavelength of the infrared absorbed is almost unique to each substance, qualitative analysis of the substance is possible by comparing it with the standard sample of infrared absorption spectrum.Past simulation test: a factory-retained traxcess was once inserted into an attached metal inserter, and then held curved and pulled out in order that the shaft section would be exposed to abrasion load by being in contact with the distal edge of metal inserter.Peeling of ptfe coat occurred in the shaft section.The condition of the test sample was considered to be similar to that of the actual sample.In the manufacturing process, the following operations and inspections are performed to maintain quality.After the coating of outer layer, 100% magnifying inspection of the shaft is performed to confirm that there is no anomaly in the appearance such as peeling of outer layer.At the time of shipping inspection, magnifying inspection is performed per lot number basis to assure that there is no anomaly in the appearance such as peeling of outer layer.Conclusion: evaluation on the return device showed a blue thread-like substance was found adhering to the distal end of the mandrel.Ftir testing was performed on the substance, and it was found to be most consistent with ptfe coating, which was found to be peeling off of the returned guidewire shaft.Based on the investigation result, no anomaly was found in the manufacturing record, the shipping inspection record, and dimensions of the actual sample.From the condition of the actual sample and the result of the simulation test, as one of the possibilities, it was inferred that the surface of shaft section came into contact with some kind of hard object (such as the tip of a metal inserter) and was exposed to abrasion load, which resulted in the peeling of ptfe coat.
 
Event or Problem Description
No additional information received.This report represents corrections to a previously submitted mdr.Fields corrected (udi related data quality updates ) are listed in h11.
 
Additional Manufacturer Narrative
Udi related data quality updates only.Upon review of device registrations and mdrs submitted to the fda, discrepancies were noted.Clarification was received regarding the device registration and the following fields have been corrected: d1, d2a, d2b, d3, d4, g5.The corrected information does not have any impact on the incident or investigative data submitted on earlier reports.The h6 codes and h11 investigation results stand as previously submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAXCESS GUIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
MICROVENTION INC
35 enterprise
aliso viejo CA 92656
Manufacturer (Section G)
MICROVENTION, INC
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18193873
Report Number2032493-2023-01071
Device Sequence Number15401394
Product Code DQX
UDI-Device Identifier00810170012280
UDI-Public(01)00810170012280(11)000101(17)260131(10)230228
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K133725
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberGW1420040
Device Lot Number230228
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2023
Supplement Date Received by Manufacturer02/06/2024
07/05/2024
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date02/08/2024
07/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2000
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-