|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Please see updates to b5.The device was returned and evaluated and found the clip had detached from the distal portion.The hook was able to extend from the distal end of the coil sheath when the slider was fully pushed.No deformation was observed in the connecting portion of the hook.The limiter in the control section of the clipping device was broken.The coil sheath was observed within specifications.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the likely cause of the reported event is due to the following: 1.When pulling the slider, the operation wire, the hook and the clip arms are linked together, and they are pulled together in the proximal direction.This motion closes the clip arms.2.When the slider is further pulled after closing the clip arms, the protrusions of the clip arms will move to the outside of the clip pipe.Once it happens, the small protrusions of the clip arms will be fix to the edge of the clip pipe.As a result, the clip cannot be opened or closed.(see fig.5) the clip can be opened and closed by operating the slider to a position where the small protrusions of the clip arms do not come out of the clip pipe.3.When the slider is further pulled after the clip does not open or close, the limiter located in the slider breaks with the noise of snapping.4.After the limiter breaks, the joint between the hook and the clip arms will be unhooked by moving the slider forward.As a result, the clip will be released from the product.The event can be detected and prevented by following the instructions for use (ifu) which state: operation of this instrument is based on the assumption that open surgery is possible as an emergency measure if the clip cannot be detached from the instrument or if any other unexpected circumstance takes place.In this case, refer to chapter 14, ¿emergency treatment¿.·do not withdraw the instrument forcibly or change the angulation of the endoscope when the clip is grasping the tissue and is not released.Doing so may tear tissue inside the body cavity, resulting in patient injury, such as perforation, bleeding, or mucous membrane damage.Do not angulate the bending section of the endoscope or withdraw the instrument from the endoscope while the clip is grasping the tissue before being released.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.¿do not try to forcibly withdraw the instrument from the endoscope if the clip cannot be detached from the instrument.Forcibly withdrawing the instrument could cause patient injury such as perforation, bleeding, or mucous membrane damage.Should the slightest irregularity and/or breakage of the parts be suspected, withdraw the instrument from the endoscope immediately according to the steps of ¿withdrawing the instrument from the endoscope¿ on page 10.Otherwise, perforation, bleeding, or mucous membrane damage may result.Olympus will continue to monitor field performance for this device.
|