• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OPTICROSS 6; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OPTICROSS 6; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 8655
Medical Device Problem Codes Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis.Visual and microscopic inspection revealed the imaging window and imaging core were twisted.Impedance testing showed an electrical open at proximal catheter and imaging core windup with a broken driveshaft beginning 26.7 cm from the distal end of the connector shaft.
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6)2023.It was reported that visualization issues occurred.The opticross hd imaging catheter was advanced for ultrasound examination of the target lesion, but the image was completely black after multiple flushes and reconnections.The procedure was completed with another of the same device.There were no patient complications reported.However, device analysis revealed the imaging window and imaging core were twisted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS 6
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18193877
Report Number2124215-2023-65646
Device Sequence Number7790703
Product Code OBJ
UDI-Device Identifier08714729904762
UDI-Public08714729904762
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K161125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/06/2025
Device Model Number8655
Device Catalogue Number8655
Device Lot Number0030995818
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/13/2023
Initial Report FDA Received Date11/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-