| Model Number |
8655 |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.Visual and microscopic inspection revealed the imaging window and imaging core were twisted.Impedance testing showed an electrical open at proximal catheter and imaging core windup with a broken driveshaft beginning 26.7 cm from the distal end of the connector shaft.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6)2023.It was reported that visualization issues occurred.The opticross hd imaging catheter was advanced for ultrasound examination of the target lesion, but the image was completely black after multiple flushes and reconnections.The procedure was completed with another of the same device.There were no patient complications reported.However, device analysis revealed the imaging window and imaging core were twisted.
|
| |
|
Search Alerts/Recalls
|