ARGON MEDICAL DEVICES BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 16GA X 10CM Back to Search Results
Model Number 370-1080-02
Medical Device Problem Code Failure to Obtain Sample (2533)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Event or Problem Description
2 subsequent ultrasound guided liver biopsies had biopsy samples taken with needle appropriately positioned, needle gun deployed, but no biopsy sample obtained.Same technique as hundreds of successful prior standard liver biopsies.This creates additional patient risk of bleeding as further biopsy samples are required.
 
Additional Manufacturer Narrative
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.Three unopened sample was returned from the customer for review.A visual inspection and testing was performed on the ret urned product.Laboratory kidney biopsy testing was performed with the three unused biopince device and it was able to obtain 9 out of 9 samples with no defects.No corrective action can be taken at this time since the issue cannot be duplicated.
 
Event or Problem Description
2 subsequent ultrasound guided liver biopsies had biopsy samples taken with needle appropriately positioned, needle gun deployed, but no biopsy sample obtained.Same technique as hundreds of successful prior standard liver biopsies.This creates additional patient risk of bleeding as further biopsy samples are required.
 
Additional Manufacturer Narrative
Udi related data quality updates only corrected information provided in d.4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPINCE ULTRA FULL CORE BIOPSY INSTRUMENT 16GA X 10CM
Common Device Name
BIOPINCE
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key18194006
Report Number0001625425-2023-01244
Device Sequence Number7790716
Product Code KNW
UDI-Device Identifier00886333224067
UDI-Public(01)00886333224067(17)260620(10)11484188
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K904987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/20/2026
Device Model Number370-1080-02
Device Lot Number11484188
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2023
Supplement Date Received by Manufacturer11/06/2023
09/04/2024
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date12/17/2023
09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-