| Model Number |
370-1080-02 |
| Medical Device Problem Code |
Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
| |
|
Event or Problem Description
|
|
2 subsequent ultrasound guided liver biopsies had biopsy samples taken with needle appropriately positioned, needle gun deployed, but no biopsy sample obtained.Same technique as hundreds of successful prior standard liver biopsies.This creates additional patient risk of bleeding as further biopsy samples are required.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.Three unopened sample was returned from the customer for review.A visual inspection and testing was performed on the ret urned product.Laboratory kidney biopsy testing was performed with the three unused biopince device and it was able to obtain 9 out of 9 samples with no defects.No corrective action can be taken at this time since the issue cannot be duplicated.
|
| |
|
Event or Problem Description
|
|
2 subsequent ultrasound guided liver biopsies had biopsy samples taken with needle appropriately positioned, needle gun deployed, but no biopsy sample obtained.Same technique as hundreds of successful prior standard liver biopsies.This creates additional patient risk of bleeding as further biopsy samples are required.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only corrected information provided in d.4.
|
| |
|