| Catalog Number |
03666794001 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The meter was requested for investigation.Replacement product was sent.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.The patient is diagnosed with antiphospholipid antibody syndrome.Per product labeling, "the presence of anti-phospholipid antibodies (apas) such as lupus antibodies (la) may lead to prolonged clotting times, i.E., they may cause false-high inr values." on a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable results from coaguchek xs meter serial number: (b)(6) compared to a laboratory result using an unknown method.The meter result was 1.8 inr.The meter result within 4 hours was 2.4 inr.No medication changes were made based on the meter results.The patient¿s therapeutic range is 2.5 - 3.0inr.The patient's testing frequency is weekly.
|
| |
|
Additional Manufacturer Narrative
|
|
The imdrf coding in this case has been updated.
|
| |
|
Search Alerts/Recalls
|