| Catalog Number |
03289923190 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received questionable results for four samples from the same patient tested with co2-l (bicarbonate) on a cobas 8000 cobas c 502 module.The results obtained for the patient did not fit the patient's diagnosis of acidosis.The first sample resulted in a bicarbonate value of 4 mmol/l on (b)(6) 2023.The second sample resulted in a bicarbonate value of 4 mmol/l on (b)(6) 2023.The third sample resulted in a bicarbonate value of 4 mmol/l on (b)(6) 2023.The fourth sample resulted in a bicarbonate value of 6.1 mmol/l on (b)(6) 2023.On an unknown date, a sample from the patient was tested on an unknown blood gas analyzer, resulting in the following values: ph = 7.35, paco2 = 49 (no units provided), pao2 = 88 (no units provided), bicarbonate = 26 mmol/l, lactate = 2.1 (no units provided).
|
| |
|
Additional Manufacturer Narrative
|
|
The serial number of the c 502 module is (b)(6).The last calibration performed on 18-oct-2023 was ok.Quality controls recovered within range.A general reagent issue can be ruled out as calibration and controls are acceptable.
|
| |
|
Additional Manufacturer Narrative
|
|
Protein electrophoresis was performed on a sample from the patient and a peak in the gamma globulin fraction was identified.Gammopathy of the patient cannot be excluded.A general reagent issue can be ruled out as calibration and controls are acceptable.The investigation determined that the issue is consistent with monoclonal gammopathy of the patient.The investigation could not identify a product problem.Per product labeling: "in very rare cases, gammopathy, in particular type igm (waldenström¿s macroglobulinemia), may cause unreliable results.".
|
| |
|
Search Alerts/Recalls
|