| Model Number |
CLK-4 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The following non-reportable malfunctions were found during the device evaluation: chassis was deformed; two feet were missing; worn out connector; faulty pump unit; broken lamp.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
| |
|
Event or Problem Description
|
|
The subject device was returned to an olympus service center with no alleged complaint.Upon inspection and testing of the returned device, a faulty transformer was found and the spare lamp was missing.There was no patient or user injury reported due to the event.This medical device report (mdr) is being submitted to capture the reportable malfunctions found during device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the shipping history is unavailable.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that the unit would not power on due to a faulty transformer.The definitive root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer reported the unit would not power on due to faulty fuse.The issue was discovered at set-up and no impact to patient or procedure.
|
| |
|
Search Alerts/Recalls
|