MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLK-4
Medical Device Problem Codes Use of Device Problem (1670); Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following non-reportable malfunctions were found during the device evaluation: chassis was deformed; two feet were missing; worn out connector; faulty pump unit; broken lamp.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
The subject device was returned to an olympus service center with no alleged complaint.Upon inspection and testing of the returned device, a faulty transformer was found and the spare lamp was missing.There was no patient or user injury reported due to the event.This medical device report (mdr) is being submitted to capture the reportable malfunctions found during device evaluation.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the shipping history is unavailable.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely that the unit would not power on due to a faulty transformer.The definitive root cause was unable to be determined.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer reported the unit would not power on due to faulty fuse.The issue was discovered at set-up and no impact to patient or procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Common Device Name
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18194585
Report Number3002808148-2023-13239
Device Sequence Number6365237
Product Code FAJ
UDI-Device Identifier04953170024511
UDI-Public04953170024511
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K993041
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2023
Supplement Date Received by Manufacturer11/24/2023
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-