• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INBONE TALAR DOME SZ 2 SULCUS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INBONE TALAR DOME SZ 2 SULCUS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 220220902
Medical Device Problem Codes Loss of Osseointegration (2408); Insufficient Device Problem Information (3190); Migration (4003)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Insufficient Information (4580)
Date of Event 10/31/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.H3 other text: device remains implanted in patient.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery of both tibial and talar components for reasons that are not available at the time of this report.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery of both tibial and talar components for reasons that are not available at the time of this report.
 
Additional Manufacturer Narrative
The reported event could be confirmed, based on available ct scans and medical expert opinion.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed: tibial components are intact.Some radiolucency around the mid stem and base stem components, not around the top stem.No signs of gross wear of pe observed.Talar component has subsided.Ct scan show overall poor bone quality after subtalar arthrodesis (fusion).Based on investigation, the root cause was attributed to a patient related issue.The failure was caused by poor bone stock that contributed to the lack of osteointegration of the talar construct resulting in subsidence.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE TALAR DOME SZ 2 SULCUS
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18196505
Report Number3010667733-2023-00715
Device Sequence Number6403334
Product Code HSN
UDI-Device Identifier00840420124210
UDI-Public00840420124210
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K100886
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number220220902
Device Lot Number1677483
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/31/2023
Supplement Date Received by Manufacturer02/05/2024
Initial Report FDA Received Date11/22/2023
Supplement Report FDA Received Date02/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-