| Catalog Number |
33680001 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
09/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented in the clinic with lucency around tibial hardware and concern for hardware malfunction.The nurse practitioner requested a ct scan from the digital stryker prophecy team indicating that the patient may require a revision surgery.The patient is treated with medication (unknown), cushioned athletic shoe, power step arch support, ice therapy, cryotherapy and voltaren gel and brace.
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.H3 other text : device remains implanted in patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented in the clinic with lucency around tibial hardware and concern for hardware malfunction.The nurse practitioner requested a ct scan from the digital stryker prophecy team indicating that the patient may require a revision surgery.The patient is treated with medication (unknown), cushioned athletic shoe, power step arch support, ice therapy, cryotherapy and voltaren gel and brace.
|
| |
|