COVIDIEN MFG DC BOULDER SONICISION; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number SCDA26
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that the jaw liner was damaged and detached.It was reported that the jaw liner of the device detached.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not to activate the instrument with the clamping jaw closed unless there is tissue present, as doing so may damage the device jaws.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the jaw broke during use on a patient.The photo observation showed that the jaw liner of the device was detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICISION
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18196830
Report Number1717344-2023-01445
Device Sequence Number6403402
Product Code LFL
UDI-Device Identifier10884521592834
UDI-Public10884521592834
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K180149
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCDA26
Device Catalogue NumberSCDA26
Device Lot Number31450047X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2023
Initial Report FDA Received Date11/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-